DRO/ORO Calibration Standard on Rxi®-5HT (15 m x 0.25 mm x 0.10 µm)

  Peaks tR (sec)
1. C10 124.3
2. C12 203.5
3. C14 279.8
4. C16 348.9
5. C18 411.5
GC_EV1380
Column Rxi®-5HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13905)
Sample DRO/ORO calibration standard (cat.# 31831)
Diluent: Methylene chloride
Conc.: 25,000 ng/µL
Injection
Inj. Vol.: 1 µL split (split ratio 50:1)
Liner: 4 mm Precision® liner w/wool (cat.# 23305.5)
Inj. Temp.: 275 °C
Split Vent Flow Rate: 87.5 mL/min
Oven
Oven Temp.: 40 °C (hold 0.1 min) to 400 °C at 19.6 °C/min (hold 1.53 min)
Carrier Gas H2, constant flow
Flow Rate: 1.75 mL/min
Detector FID @ 420 °C
Constant Column + Constant Make-up: 50 mL/min
Make-up Gas Type: N2
Hydrogen flow: 40 mL/min
Air flow: 450 mL/min
Data Rate: 20 Hz
Instrument Agilent/HP6890 GC
Notes Actual column length was calculated to be 15.7 m using holdup time and EZGC® flow calculator (http://www.restek.com/ezgc-mtfc).

 